Model 806 HAADF STEM Detector
Model 806 HAADF STEM Detector

Figure 1 New Gatan 806 HAADF STEM detector head.
Specification
Key Benefits
1. Optimized Geometry
For EELS analysis, the correct choice of camera length and EELS collection
angles is critical. The same is true for Z-contrast (also referred as high angle
angular dark field – HAADF) STEM imaging. The Gatan model 806 HAADF STEM
detector is specifically designed to capture the optimum range of scattering
angles for Z-contrast STEM at camera lengths optimally suited to simultaneous
EELS acquisition with any Gatan EELS system (GIF2000, Tridiem or Enfina). The
model 806 mounts to a Gatan UltraScan camera housing and is compatible with
any 100-300 kV TEM/STEM that can accommodate this housing. Model 806 can be
mounted together with the model 894/895 UltraScan or model 832 ORIUS CCD cameras.
The detector geometry and mounting position are compatible with concurrent deployment
of Gatan’s model 805 BF/DF STEM detector, thus enabling simultaneous and
optimized EELS, Z-contrast and diffraction-contrast STEM imaging when combined
with Gatan model 777 STEMPack.
2. Single electron detection
The 806 detector and its electronics have been optimized for the low to moderate
count rates typically encountered in HAADF STEM work. The system has very low
noise and is capable of detecting single electron events, both in analog and
pulse-counting modes.

Figure 2 Histogram of analog mode pixel charges.
Figure 3 Pulse Height Analysis using the 806 HAADF Detector, showing the
single electron detection capabilities of the system.
The data were gathered using a 200 KeV STEM, biasing the Photomultiplier tube
at -760 V. The discriminator threshold was swept from pulse amplitudes of zero
to 1.8 electron charges. The noise distribution has a Standard Deviation of
0.06 e- , and the gain distribution has a Standard Deviation of 0.25 e-.
3. Calibrated output
When integrated as part of a complete STEMPack configuration (with DigiScan
II), the model 806 HAADF detector is automatically calibrated in units of primary
electrons detected. This enables true quantitative interpretation of Z-contrast
STEM images and the counting statistics of the signal. Installation with Gatan’s
original 688 DigiScan or a 3rd party digital STEM system yields the same outstanding
low-noise, computer controlled operation but with the output in arbitrary digitizer
units.
4. Ease of use
The calibration of the detector greatly simplifies its operation since the
zero-count baseline (black level) is always known and fixed, thereby requiring
only the adjustment of the detector gain as the beam intensity varies. The detector
gain (PMT voltage) is automatically and quickly optimized with a single mouse
click. For operators who prefer manual control, direct digital gain control
is also provided. The detector output remains calibrated in true electron counts
over the entire range of practical gain and dwell time settings.
5. Hardware configuration
The electronics of the system have been unified into a single, 19” chassis
that is a standard part of the product package. Detector insertion and operation
settings are all software controlled within DigitalMicrograph, the primary acquisition,
control, and data handling application in the Gatan Microscopy Suite (GMS) software
system. There is no need for manual access to the detector electronics, which
can thus be mounted in a remote rack well away from the TEM column. While the
806 HAADF detector in principle can be installed as a stand-alone STEM detector,
it is strongly recommended that the system be installed together with STEMPack
or, at minimum, a DigiScan II to realize the full capabilities and ease of use
of the system.
Application examples
1. Z-contrast vs. diffraction contrast in STEM imaging
Z-contrast STEM images are formed by collecting electrons scattered to high
angles (~50 to 200mrad) where the scattering is predominately incoherent and
roughly proportional to the square of the atomic number, Z, of the scattering
atom. This contrast mechanism can yield images where even small changes in average
number are easily discriminated. Moreover, the incoherent nature of the scattering
eliminates the contrast reversals that can dominate coherent scattering images
and can provide a 1.5x increase in spatial resolution over coherent imaging.
On the other hand, conventional angular dark-field (ADF) images formed using
electrons scattered low to medium angles (~10 to 50mrad) show strong diffraction
contrast. Diffraction contrast is useful in studying the crystalline nature
of the specimen, for example grain size, defects, strain fields, etc.
The 806 HAADF and 805 BF/DF detectors are designed to complement each other,
and work together with Gatan EELS analysis equipment. The carefully designed
placement and size of the two detectors allow the simultaneous and optimized
acquisition EELS data together with HAADF and ADF STEM resulting in a complete
picture of the sample composition and structure.
| |
|
| |
|
Figure 4 STEM images of Al-Cu sample
recorded simultaneously with Gatan 806 HAADF STEM detector (left) and
805 BF/DF STEM detector (right). The DF image shows strain fields and
defects in the sample, but the smallest precipitates are difficult to
identify. In the HAADF image, the diffraction contrast is greatly minimized
giving a clear accounting of the precipitates. |
| |
|
| |
|
Figure 5 STEM image of a semiconductor
device recorded simultaneously with Gatan 806 HAADF STEM detector (left)
and 805 BF/DF STEM detector (right). The Z-contrast effect in the HAADF
image allows unambiguous identification of the material layers, while
the low-angle ADF image is difficult to interpret due to strain and
contrast reversal effects. However, the low-angle ADF image shows fine
structures not evident in the HAADF image. |
2. The optimized STEM solution - Putting it all together!
Model 806 HAADF STEM detector is a great addition to STEMPack and EELS analysis.
This integrated and optimized system provides efficient access to all possible
STEM-mode signals within Gatan’s GMS software environment. Hence it allows
users to gain a complete understanding of the structure and chemistry of the
sample under investigation.

Ordering information
Model No |
Description
|
|
806 |
HAADF STEM detector system for up to 300 keV beam energy – includes
detector head assembly, controller, and cables as well as GMS software
license for detector calibration, detector control, and DigiScan II pulse
input option
|
|
806.T |
Installation and training for Model 806 HAADF STEM detector |
For a PDF version of the 806 HAADF STEM detector web page, please click here.
|